1 electrical connection, 1 power cord – HEIDENHAIN MSE 1000 User Manual
Page 25

Installation
Electrical connection
Operating Instructions
25
5.1
Electrical connection
WARNING
A risk of electrical shock exists if this product is not properly grounded.
To avoid the hazard, always use a 3-conductor (grounded) power cord and ensure the
ground is properly wired to the building installation.
WARNING
A risk of fire exists if a power cord not meeting minimum specifications is used.
To avoid the hazard, always use power cords that meet or exceed the minimum
listed specifications.
5.1.1
Power cord
MSE 1201
(ID 747501-01)
MSE 1201
(ID 747501-02)
MSE 1202
(ID 747502-01)
MSE 1201
MSE 1202
L: Line voltage (brown)
1
DC 24 V
N: Neutral (blue)
3
0 V
Protective conductor (ground)
terminal(green/yellow)
4
Protective conductor (ground)
terminal
3-conductor (grounded)
Min. wire cross section: 0.24 mm
2
Min. cable cross section: 0.75 mm
2
Max. wire cross section: 0.35 mm
2
Max. cable length: 3 m