Troubleshooting
Application errors
Operating Instructions
141
Error
Cause
Corrective action
Duplicate IP Address
More than one module has the
same IP address.
Clearing the error will cause the
error to go away but any resulting
module chain will not be usable until
the problem is corrected.
DHCP addressing:
Restore factory defaults.
Static addressing:
Uninstall one of the modules with
the duplicate IP address from the
module chain.
Change the IP address of the other
module to a unique value.
Reinstall the removed module.
Error
A Query failed.
A Communication error and the
IP address of the module will
displayed.
Power cycle the module chain.
Refer to "Power cycling", page 33.
Perform a Broadcast.
Refer to "Network troubleshooting",
Could not create the MSE 1000
module chain via broadcasting
A Broadcast failed.
Verify the network connection is
connected.
Verify power to the modules is on.
Verify the workstation IP address.
Verify the workstation netmask.
Verify the broadcast netmask.
Verify the workstation port.
Power cycle the module chain.
Refer to "Power cycling", page 33.
Wait for the Network bus LED to
flash 2/sec on all modules.
Perform a Broadcast.
Refer to "Network troubleshooting",