Initial power-up

Power cycling

Operating Instructions

33

6

Initial power-up

MSE 1201
(ID 747501-01)

The MSE 1201 (ID 747501-01) power supply module has a power switch

5

for turning

power on and off.

To power-up the system:

Press the On (supply) side of the power switch to power-up the system.

MSE 1201
(ID 747501-02),

MSE 1202
(ID 747502-01)

The MSE 1201 (ID 747501-02) and MSE 1202 (ID 747502-01) power supply modules do
not have a power switch for turning power on and off.

To power-up the system:

Turn power on for the system.

6.1

Power cycling

Power cycling refers to the procedure of turning the power provided to the MSE 1000
modules off and then back on. Power cycling is typically used when troubleshooting
communication problems between MSEsetup and the modules and may allow the
modules to reinitialize their configuration or recover from an unresponsive state.

To power cycle the module chain:

Power-off the module chain. Refer to "Power-off", page 155.

Wait twenty seconds.

Power-up the module chain. Refer to "Initial power-up", page 33.