8 commissioning and preparing to measure, 1 network configuration, 1 connect screen – HEIDENHAIN MSE 1000 User Manual
Page 63: Commissioning and preparing to measure, 8commissioning and preparing to measure

Commissioning and preparing to measure
Network configuration
Operating Instructions
63
8
Commissioning and preparing to measure
8.1
Network configuration
MSEsetup communicates with MSE 1000 modules over Ethernet. The PC running
MSEsetup is referred to as the Workstation and is the Client when communicating with
the modules.
The workstation must have a Network Interface Controller (NIC) with a unique Internet
Protocol (IP) address that is on the same subnetwork as the MSE 1000 modules. All
MSE 1000 modules must also have a unique IP address.
Default IP addresses for similar module types are set the same from the factory and
need to be reconfigured in order to communicate with MSEsetup. Workstation and
MSE 1000 module IP address configurations and other network settings and operations
are located in the Connect screen.
Model
Default IP
address
Default netmask
MSE 120x
172.31.46.2
MSE 11xx, MSE 13xx, MSE 1401, MSE 1501
172.31.46.1
255.255.255.0
8.1.1
Connect screen
The Connect screen is available without a password when MSEsetup is opened for
the first time. All subsequent MSEsetup sessions require enabling Supervisor mode to
access the Connect screen.
Available options:
Workstation IP
IP address
Netmask
Port
Async. Port
Individual Module IP
Query
Set Static
Connection
Disconnect
Reconnect
Broadcast
Module chain
Use DHCP Addressing
Use Static Addressing
Restore Factory Defaults