Commissioning and preparing to measure

Network configuration

Operating Instructions

63

8

Commissioning and preparing to measure

8.1

Network configuration

MSEsetup communicates with MSE 1000 modules over Ethernet. The PC running
MSEsetup is referred to as the Workstation and is the Client when communicating with
the modules.

The workstation must have a Network Interface Controller (NIC) with a unique Internet
Protocol (IP) address that is on the same subnetwork as the MSE 1000 modules. All
MSE 1000 modules must also have a unique IP address.

Default IP addresses for similar module types are set the same from the factory and
need to be reconfigured in order to communicate with MSEsetup. Workstation and
MSE 1000 module IP address configurations and other network settings and operations
are located in the Connect screen.

Model

Default IP
address

Default netmask

MSE 120x

172.31.46.2

MSE 11xx, MSE 13xx, MSE 1401, MSE 1501

172.31.46.1

255.255.255.0

8.1.1

Connect screen

The Connect screen is available without a password when MSEsetup is opened for
the first time. All subsequent MSEsetup sessions require enabling Supervisor mode to
access the Connect screen.

Available options:

Workstation IP

IP address

Netmask

Port

Async. Port

Individual Module IP

Query

Set Static

Connection

Disconnect

Reconnect

Broadcast

Module chain

Use DHCP Addressing

Use Static Addressing

Restore Factory Defaults