6 firewall configuration – HEIDENHAIN MSE 1000 User Manual

Page 72

Advertising
background image

Commissioning and preparing to measure

Network configuration

72

Operating Instructions

To restore factory defaults:

Click the «Restore Factory Defaults» button.

A dialog box requests confirmation to restore the modules to factory default
settings.

Click the «OK» button.

Cycle power for the entire module chain by turning the power off and then back on.

The Network LED on each module will flash five times per second while waiting for
DHCP to assign an IP address. The Network LED on each module will flash twice
per second when an address has been assigned.

If a router with a DHCP server server is not used the modules need to be manually
programmed to unique static addresses. Refer to "Configuring the module chain
manually", page 75.

Wait for the Network LED to flash twice per second on all modules.

Click the «OK» button in the dialog box.

MSEsetup performs a broadcast. Each modules Network LED will flash twice per
second until MSEsetup obtains a connection to the module. A dialog box confirms
the number of modules found.

Click the «OK» button in the dialog box to overwrite the configurable module
settings with a backup file or the «Cancel» button to skip overwriting the the
settings.

A message is displayed in the Connect screen message area when broadcasting is
finished.

8.1.6

Firewall configuration

MSEsetup uses the UDP protocol for communication with MSE 1000 modules. By
default firewalls may block UDP messages. Network firewall settings may need
to be configured to allow UDP on the ports used by MSEsetup and MSE 1000
modules. Refer to the documentation provided with the firewall software for additional
information.

MSE port assignments:

MSE component

Port

MSEsetup

27016

The port can be changed in the Connect
screen. Refer to "Workstation IP", page 65.

MSE 1000 modules

27015

This port cannot be changed.

MSEsetup

asynchronous

thread

27300

The port can be changed in the Connect
screen. Refer to "Workstation IP", page 65.