Commissioning and preparing to measure

Module setup

Operating Instructions

81

8.2.2

LVDT Teach resolutions

The LVDT Teach resolutions screen is used to convert the voltage value returned from
an LVDT sensor into millimeters or inches. The conversion is done by interpolating
the voltage based on the minimum and maximum physical position of the sensor. The
Teach resolutions screen configures all of the sensors connected to a single LVDT
module. LVDT sensors can be configured individually from the Setup screen channel
and device tabs. Teach resolutions only applies to a channel that has a sensor assigned
to it.

Teach resolutions information:

Min (mm) or Min (in)

Max (mm) or Max (in)

Sensor output (V)

Res. (mm/V) or Res. (in/V)

Calibration timestamp

LVDT Teach
resolutions screen