Installation

Data interface connections

28

Operating Instructions

5.2.2

Connecting a network cable

15

RJ-45 connector

MSE modules communicate with a PC workstation through an RJ-45 connection

15

located on the base module. Use a straight-through cable when connecting to a DHCP
router. A crossover cable may be required when connecting the workstation directly
to the base module. Refer to the documentation provided with the NIC for additional
information.

To connect a network cable:

Verify power is removed from all power supply modules in the module chain.

With the cable connector locking tab facing up, insert the cable connector into the
RJ-45 connection

15

on the module until the tab locks into place.

5.2.3

Connecting an EnDat encoder

16

EnDat encoder
connector

This product can be used with HEIDENHAIN length gauges, linear encoders and rotary
encoders that provide EnDat signals. The connecting cable must not exceed 100
meters in length.

To connect an EnDat encoder:

Verify power is removed from all power supply modules in the module chain.

Align the encoder cable connector notch with the module encoder connection
notch.

Insert the cable connector into the encoder connection

16

and tighten the cable

connector by turning clockwise until it is snug.