Super Systems 3 Series User Manual

Page 76

Advertising
background image

Operations Manual

Series 3

76

Parameter
Mnemonic

Parameter Name

Modbus Address

Decimal

PNT.LO

Adjust Low Point

563

CT.RNG

CT Range

572

Sb.tyP


Sensor Break Type
0 – No Sensor Break
1 – Non-Latching Sensor Break
2 – Latching Sensor Break

578


Id

Customer ID – May be set to any value between 0-9999 for identification of instruments in applications. Not used by the
instrument itself.

629

PHASE








Calibration Phase
0 – None
1 – 0 mv
2 – 50 mv
3 – 150 Ohm
4 – 400 Ohm
5 – CJC
6 – CT 0 mA
7 – CT 70 mA
8 – Factory Defaults
9 – Output 1 mA low cal
10 – Output 1 mA high cal
11 – Output 2 mA low cal
12 – Output 2 mA high cal
13 – Output 3 ma low cal
14 – Output 3 ma high cal
15 – Remote setpoint input low volts
16 - Remote setpoint input high volts
17 - Remote setpoint input low current
18 - Remote setpoint input high current

768








GO

Calibration Start
0 – No
1 – Yes (start cal)
2 – Cal Busy
3 – Cal Pass
4 – Cal Fail
Note values 2-4 cannot be written but are status returns only

769

-

Analogue Output Calibration Value

775

K.LOC

Allows instrument to be locked via a key/digital input
0 - unlocked,
1 – all keys locked
2 – Edit keys (raise and lower) disabled
3 – Mode key disabled
4 – Manual mode disabled
5 – Enter standby mode when Mode combination pressed

1104

Dwel.1

FEATURE NOT AVAILABLE

1280

TSP.1

FEATURE NOT AVAILABLE 1

1281

RMP.1

FEATURE NOT AVAILABLE

1282

Dwel.2

FEATURE NOT AVAILABLE

1283

TSP.2

FEATURE NOT AVAILABLE

1284

RMP.2

FEATURE NOT AVAILABLE

1285

Dwel.3

FEATURE NOT AVAILABLE

1286

TSP.3

FEATURE NOT AVAILABLE 3

1287

RMP.3

FEATURE NOT AVAILABLE 3

1288

Dwel.4

FEATURE NOT AVAILABLE

1289

TSP.4

FEATURE NOT AVAILABLE 4

1290

RMP.4

FEATURE NOT AVAILABLE

1291

AT.R2G

Auto-tune Configures R2G
0 - YES
1 - No

4176

IN.TYP

Input Sensor Type
0 – J Type Thermocouple
1 – K Type Thermocouple
2 – L Type Thermocouple
3 – R Type Thermocouple

12290