To Calibrate Thermocouple Input

Thermocouples are calibrated, firstly, by following the previous procedure for the mV ranges, then calibrating the CJC.
Connect a mV source. Set the mV source to ‘internal compensation’ for the thermocouple in use and set the output for 0mV.
Then

Operation

Do This

Display View

Additional Notes

Select the Calibration List
header

1.

From any display press

as many

times as necessary until the ‘C A L ’ page header
is displayed.

Select the calibration phase

2.

Press

to select ‘P H A S E ’

n o n e

p h a s e

Scrolling display ‘

C A L I B R A T I O N

p h a s e

Select CJC calibration

3.

Press

or

to select ‘CJC’

Calibrate CJC

4.

Press

to select ‘GO’

5.

Press

or

to choose ‘YES’

The controller automatically calibrates to
the CJC input at 0mV. The display will show

busy

then

pass

, (if calibration is successful)

or ‘

FAIL’

if not. Fail may be due to an

incorrect input mV