3 to calibrate rtd input, To calibrate rtd input, N o n e – Super Systems 3 Series User Manual

Page 86

Advertising
background image

Operations Manual

Series 3

86

14.3.3

To Calibrate RTD Input

The two points at which the RTD range is calibrated are 150.00Ω and 400.00Ω.
Before starting RTD calibration:

A decade box with total resistance lower than 1K must be connected in place of the RTD as indicated on the connection
diagram in section 15.1.4 before the instrument is powered up. If at any time the instrument was powered up without this
connection then at least 10 minutes must elapse from the time of restoring this connection before RTD calibration can take
place.

The instrument should be powered up for at least 10 minutes.

Before calibrating the RTD input the mV range must be calibrated first

Operation

Do This

Display View

Additional Notes

Select the Calibration List
header

1.

From any display press

as many

times as necessary until the ‘C A L ’ page header
is displayed.

Scrolling display ‘

C A L I B R A T I O N

L I S T

Select the calibration phase

2.

Press

to select ‘P H A S E ’

n o n e

p h a s e

Scrolling display ‘

C A L I B R A T I O N

p h a s e

Set the decade box for 150.00

Select the low calibration
point (150Ω)

3.

Press

or

to choose ‘150r

Calibrate the low point

4.

Press

to select ‘GO’

5.

Press

or

to choose ‘YES’

Scrolling display ‘

C A L I B R A T I O N

s t a r t

The controller automatically calibrates to the injected 150.00Ω input. The display will show

busy

then

pass

(if calibration is successful) or ‘

FAIL’

if not. Fail may be due

to an incorrect input resistance

Set the decade box for 400.00

Select the high calibration
point (400Ω)

7.

Press

or

to choose ‘400r

Calibrate the high point

8.

Repeat 5 and 6 above to calibrate the high

point

The controller will again automatically calibrate to the injected 400.00Ω input. If it is not successful then ‘

FAIL’

will be displayed

Advertising