4 to calibrate ma outputs, To calibrate ma outputs – Super Systems 3 Series User Manual

Page 87

Advertising
background image

Series 3

Operations Manual

87

14.3.4

To Calibrate mA Outputs

Output 3 is supplied as mA outputs. The output may be adjusted as follows
Connect an ammeter to the output – terminals 3A/3B as appropriate.







Then, in configuration level

Operation

Do This

Display View

Additional Notes

Select low point calibration phase for
the mA output to be calibrated (eg
OP3)

1.

From the ‘CAL’ list header press

to select

‘PHASE’

2.

Press

or

to choose 1ma.L’

Scrolling message ‘calibration
phase

Set the low point output

3.

Press

to select ‘V A L U E ’

4.

Press

or

to adjust this value so that it

reads the same value as shown on the ammeter.
For example if the meter reads 2.06 then set the
controller reading for 206. The decimal point is
not displayed on the controller so that 200
represents 2.00.

Scrolling message ‘dc output
reading

Select high point calibration phase
for the mA output to be calibrated
(eg OP3)

5.

Press

to go back to ‘PHASE’

6.

Press

or

to choose 1ma.H’

Scrolling message ‘calibration
phase

Set the high point output

7.

Press

to select ‘V A L U E ’

8.

Press

or

to adjust this value so that it

reads the same value as shown on the ammeter.
The value represents 18.00mA

Scrolling message ‘dc output
reading


Controller

A +ve

B -ve

2.00

Digital ammeter

Advertising