2 offsets, 1 two point offset, Offsets – Super Systems 3 Series User Manual

Page 82: Two point offset

Advertising
background image

Operations Manual

Series 3

82




Figure -3: Connections for RTD Calibration


The RTD range of the instrument is -200 to 850

O

Set the resistance of the decade box to the minimum range.
For example 0

C. It is,

however, unlikely that it will be necessary to check the
instrument over this full range.

O

Set the resistance of the decade box to the maximum range.
For example 200

C = 100.00Ω. Check the calibration is within

+0.25% of reading + 1LSD.

O

14.2 Offsets

C = 175.86Ω. Check the calibration is

within +0.25% of reading + 1LSD.

The process value can be offset to take into account known
errors within the process. The offset can be applied to any
Input Type (mV, V, mA, thermocouple or RTD).
A single offset can be applied - the procedure is carried out in
the I NP UT list..
It is also possible to adjust the low and high points as a two
point offset. This can only be done in Level 3 in the ‘Ca l
list and is described below.

14.2.1

Two Point Offset

A two point offset adjusts both a low point and a high point
and applies a straight line between them. Any readings
above and below the calibration points will be an extension
of this straight line. For this reason it is best to calibrate with
the two points as far apart as possible as shown in the
example below:























Figure 4 Two Point Offset Applied to Linear and Non-linear Inputs


Matched impedance copper

leads

Decade

Box

Controller

VI

V-

V+

Electrical
Input

Display Reading

High offset

(e.g. 10.0)

Factory
calibration

Low offset (e.g.

8.0)

Electrical
Input

Display Reading

High offset

Factory
calibration

Low offset

Advertising