2 to apply a two point offset, 3 to remove the two point offset, To apply a two point offset – Super Systems 3 Series User Manual

Page 83: To remove the two point offset, 8 .0 c .a d j, C .a d j, C a l

Advertising
background image

Series 3

Operations Manual

83

14.2.2

To Apply a Two Point Offset

Assume the instrument is set up to display 0.0 for an input of 4.00mV and 500.0 for an input of 20.00mV. Assume that a
particular sensor in use has known errors such that the instrument is required to read 8.0 for an input of 4.00mV and 490.0 for an
input of 20.00mV. To compensate for these errors in the process a low point offset of 8.0 and a high point offset of 10.0 can be
set as follows

Operation

Do This

Display View

Additional Notes

Select Calibration list header

1.

Select Level 3. Then press

to select

‘CAL’

c a l

Two pint offset can only be carried out in Level 3

Set mV input to 4.00mV

Select User Calibration

2.

Press

to scroll to ‘U.CAL’

I d L e

u c a l

Scrolling 2message

u s e r c a l i b r a t i o n

Select Low calibration point

3.

Press

or

to ‘LO’

L o

u c a l

Set the low offset value

4.

Press

to scroll to ‘C.ADJ’

5.

Press

or

to set the low offset

value eg 8.0

8 .0

c .a d j

This applies an offset over the whole range in the
same way as a simple offset.

6.

The controller then reverts to the CAL list

header

C A L

This is the same as 1 above

Set mV input to 20.00mV

Select User Calibration

7.

Press

to scroll to ‘U.CAL’

I d L e

u c a l

This is the same as 2 above

Select the high calibration
point

8.

Press

or

to ‘HI’

Hi

u c a l

Select the high calibration
offset parameter

9.

Press

to scroll to ‘C.ADJ’

5 0 8 .0

c .a d j

The reading will show 508.0

Set the high offset value

10. Press

or

to set the high offset

value to read 490.0

4 9 0 .0

c .a d j

Under normal operating conditions the controller will now read 8.0 for an input of 4.000mV and 490.0 for an input of 20.000mV.

14.2.3

To Remove the Two Point Offset

Operation

Do This

Display View

Additional Notes

In level 3 select the
Calibration list header

1.

In Level 3, press

to select ‘CAL’

c a l

Two point offset can only be carried out in Level
3

Select User Calibration

2.

Press

to scroll to ‘U.CAL’

I d L e

u c a l

Scrolling message

u s e r c a l i b r a t i o n

Reset to no offset

3.

Press

or

to select ‘r.set’

r s e t

u c a l

The display will revert to 2 above and the two point offsets will be removed.

Advertising