Transmission control for primary transmissions, Table 19, Transmission control for primary – Vaisala Hydromet Data Collection Platform User Manual
Page 177

Chapter 4 ________________________________________________ Configuring Telemetry Options
VAISALA______________________________________________________________________ 175
Examples:
The transmitter ID is set to "010012D4" and stored to EEPROM as
follows:
open
/ > CFG GOES_1 setid_eeprom "010012D4"
/ > LASTVAL GOES_1 service
Status:1 Value: ID set
/ >
The transmitter clock is set to 15.11.2005 13:17:00 as follows:
/ > CFG GOES_1 setclock "051115131700" / >
/ > LASTVAL GOES_1 service
Status:1 Value: Clock set
/ >
Transmission Control for Primary Transmissions
The parameters that are needed for controlling the transmissions are set
with the following command:
spset <parameter> <value>
You must set the parameters listed in
Table 19
Transmission Control Parameters for Primary
Transmissions
Parameter
Value
Description
goes_txtime
0...86400 seconds Time of the first transmission on
each day, given in seconds since
midnight
goes_window
0...n seconds
Length of the transmission window in
seconds
goes_interval
0...n seconds
Interval between transmissions in
seconds
goes_channel
0...266
Transmission channel number