4.5.3 XTC/3M Outputs

The Relay Outputs and TTL Outputs screens allow modification of the 12 relays
or the 8 TTL outputs, respectively. All relays (1 to 12) and TTL Open Collector
outputs (13 to 20) are selectable to perform the functions shown on the screen. Any
number of outputs may be assigned to the same function. The default assignments
are identical to the default configuration provided in XTC/2 and referenced above.

On the right side of the parameter panel will be a column of the outputs as they are
currently programmed. There will be blanks if the output is not defined.

The scrollable list of allowed output functions will be displayed on the left side of
the parameter panel. The

▲▼◄►

keys will cursor through this list. With the

cursor on <More>, press the

arrow to access additional functions. With the

cursor on <Previous>, press the

arrow to return. Entering a number (1-12 for

relays, or 13-20 for TTL), followed by the ENTER key, will assign the selected
output function to that output number. Moving the cursor to the extreme right
column permits changing the output between NO (Normally Open) and NC
(Normally Closed) by using the TOGL key. With the unit powered off, the outputs
will always be open.

The

◄►

arrow keys will go back and forth between the selection panel and the

parameter panel. For the complete list of output functions available for XTC/3M,
see

Table 3-2 on page 3-19

.

Figure 4-24 Relay Outputs 1