INFICON XTC/3 Thin Film Deposition Controller Operating Manual User Manual

Page 66

Advertising
background image

3 - 20

PN

07

4-

44

6-

P1

J

XTC/3 Operating Manual

Ready (M only)

When each State starts, the old
and new state output are set or
cleared as necessary

When each State starts, the old and
new state output are set or cleared
as necessary

Crucib Switch
(M only

Rise 1 (M only)

Soak 1 (M only)

Rise 2 (M only)

Soak 2

Shutter Delay
(M only)

Deposit (M only)

Rate Ramp (M only)

Manual (M only)

Time Power (M only)

Idle Ramp (M only)

Idle (M only)

Max Power (M only)

When power reaches max power

When power is not at max power

Rate Dev Fail
(M only)

When the rate deviation is greater
than 5% or 0.5Å of the desired
rate for 60 seconds or 20 times
the time constant if the time
constant is greater than 3
seconds

When rate deviation is within test
limits.
When Idle Ramp is entered

Xtal Sw Fail
(M only)

When an attempted crystal switch
fails

When another crystal switch is
started

Xtal Switching
(M only)

Whenever a crystal switch is
active

When the crystal switch is done

Crucib Sw Fail
(M only)

No crucible valid input within 30
seconds of a crucible switch

When Start or Reset is done.
If Stop on Alarms is set to No, XTC/3
will give the error message but will
not do anything else, it stays in
Crucible Switch. If the Crucible Valid
input becomes set, the output is
cleared

Table 3-2 Output states (continued)