3.5.1 Alarms and Stops

There are a number of unusual XTC/3 situations that may require operator
attention. These situations are detected and then treated as ALARMS or STOPS.
Both ALARMs and STOPs can be indicated by a separate relay closure (by default
or selected assignment in XTC/3M, by fixed assignment in XTC/3S).

An ALARM condition is not fatal, XTC/3 will continue the layer or process
to normal termination.

A STOP is fatal, immediately halting the process. If desired, the user may set the
STOP ON ALARM parameter in the Process sub-screen of the General
Parameter
display to configure XTC/3 to treat an ALARM the same as STOP
(i.e., halt processing upon detection of the abnormal condition).

3.5.1.1 Alarms

The following conditions are considered ALARMs by XTC/3 and result in the
closure of the ALARM relay (if assigned):

Crucible hearth selection is not validated by the CRUC VALID input
within thirty seconds.

Rate control not established during the first sixty seconds of SHUTTER DELAY
(or twenty times the TIME CONSTANT if greater).

Rate has been out of control in DEPOSIT for sixty seconds (or twenty times the
TIME CONSTANT if greater). Out of control is defined as a rate error >0.5 Å/s
and >5% of setpoint deposit rate.

The source power has exceeded the Maximum Power parameter, set in the
Source sub-screen of the Film Parameter display, for five continuous seconds.

3.5.1.2 Stops

The following actions or conditions produce a STOP state. This condition is
indicated by the STOP message on XTC/3 and the closure of the STOP relay
(if assigned):

An intentional or accidental interruption of power, indicated by
the Power Loss message.

A Switcher Fail message appears as the result of a failure to detect a valid
position in a multi-position sensor.

Pressing the front panel STOP key on XTC/3 or handheld controller.

Initiating a STOP via remote communication.

Activating the STOP external input.

A CRYSTAL FAIL detected during any pre-deposit phase (when crystal
switching is not available).