3 standard update film parameters – INFICON XTC/3 Thin Film Deposition Controller Operating Manual User Manual

Page 133

Advertising
background image

5 - 21

PN

07

4-

44

6-

P1

J

XTC/3 Operating Manual

5.2.2.4.3 Standard Update Film Parameters

Command:

UF <Command ID> <Film Number> <Parameter Value>

Command ID = <Byte>. See “UF Command ID” column of

Table 5-2

.

Film Number = <Byte> 1-9 (XTC/3S ) or 1-32 (XTC/3M)

Parameter Value = <Integer> | <Float>

Integer = 4 byte, low to high

Float = 4 byte, ANSII standard, single precision, low to high

Description: Type of parameter value depends on command ID,
see “DataType / Format” column of

Table 5-2

.

Response:

None (Just header and trailer)

Special Case: UF255 <Film Number> <List of all Film Parameter Values in
order> will update all film parameters for the given film, in numeric order.

NOTE: Some limitations apply as to when specific film parameters may be

updated, e.g., Hardware used by a film cannot be redefined while that film
is active. All parameters may be updated in Ready state.

Advertising