2 standard query film name (xtc3/m only), 3 standard query film parameters, 4 standard query general parameter – INFICON XTC/3 Thin Film Deposition Controller Operating Manual User Manual
Page 125

5 - 13
PN
07
4-
44
6-
P1
J
XTC/3 Operating Manual
5.2.2.3.2 Standard Query Film Name (XTC3/M only)
Command:
QN <Film Number>
Film Number = <Byte> 1-32
Response:
<String> (maximum 15 characters, plus terminating null character)
String = Null terminated series of ASCII characters
5.2.2.3.3 Standard Query Film Parameters
Command:
QF <Command ID> <Film Number>
Command ID = <Byte> See “UF Command ID” column in
Film Number = <Byte> 1-9 (XTC/3S ) or 1-32 (XTC/3M)
Response:
<Integer> | <Float>
Integer = 4 byte, low to high
Float = 4 byte, ANSII standard, single precision, low to high
Description: Type of response depends on command ID, see “DataType /
Format” column of
.
Special Case: QF255 <Film Number> will return all film parameters for the
given film, in numeric order.
5.2.2.3.4 Standard Query General Parameter
Command:
QG <Command ID>
Command ID = <Byte> See “UG Command ID” column in
Response:
<Integer> | <Float>
Integer = 4 byte, low to high
Float = 4 byte, ANSII standard, single precision, low to high
Description: Type of response depends on command ID, see “DataType /
Format” column of
.
Special Case: QG255 will return all general parameters, in numeric order