6.3.2 Troubleshooting Transducers/Sensors

Many sensor head problems may be diagnosed with a DVM (Digital Volt Meter).
Disconnect the short oscillator cable from the feedthrough and measure the
resistance from the feedthrough’s center pin to ground. If the reading is less than
1-2 MΩ, the source of the leakage should be found and corrected. Likewise, with
the vacuum system open check for center conductor continuity, a reading of more
than 1 Ω from the feedthrough to the transducer contact indicates a problem.
Cleaning contacts or replacing the in-vacuum cable may be required.

A more thorough diagnosis may be performed with the optional Crystal Sensor
Emulator, PN 760-601-G2. See

section 6.5 on page 6-24

for a discussion of its use

and diagnostic capabilities.

6.3.2.1 Verifying Crystal Switching and Rotation

The Sensor Information screen allows checking active sensor function from XTC/3
front panel. Refer to

section 3.3.8, Sensor Information Display, on page 3-13

.

6.3.2.2 Verifying Sensor Shutter Operation

To exercise the shutter on a Crystal Two, use the Switch Crystal function as
described under

section 3.3.8

. To exercise the shutter on a shuttered sensor,

(Standard, Compact, Sputtering, CrystalSix, Crystal12, Rotary or UHV Bakeable
sensor with shutter), use the Diagnostics screen Toggle Shutter function for the
appropriate sensor per

Figure 6-2 on page 6-6

above.

NOTE: A more detailed troubleshooting guide is shipped with the sensor.

Refer to the sensor operating manual.