4 standard update general parameters – INFICON XTC/3 Thin Film Deposition Controller Operating Manual User Manual
Page 137

5 - 25
PN
07
4-
44
6-
P1
J
XTC/3 Operating Manual
5.2.2.4.4 Standard Update General Parameters
Command:
UG <Command ID> <Parameter Value>
Command ID = <Byte> See “UG Command ID” column of
.
Parameter Value = <Integer> | <Float>
Integer = 4 byte, low to high
Float = 4 byte, ANSII standard, single precision, low to high
Description: Type of parameter value depends on command ID, see
“DataType” column of
.
Response:
None (Just header and trailer)
Special Case: UG255<List of all General Parameter Values in order> will
update all general parameters in numeric order.
NOTE: Some limitations may apply as to when general parameters
may be updated.
Table 5-3 Standard general parameters
UG
(0x5547)
Command
ID
Name
Units / Allowed
Values / Notes
Data
Type /
Format
Low
Limit
High
Limit
0
(0x00)
1
Process to run
(XTC/3M only)
Can be changed only
in Ready or Idle in last
layer
Int
1
99
1
Film to run
(XTC/3S only)
1
9
1
(0x01)
Start Layer Without
Backup Xtal
1 = Yes, 0 = No
Int
0
1
2 (0x02)
Stop on Alarms
1 = Yes, 0 = No
Int
0
1
3 (0x03)
Stop On Max Power
1 = Yes, 0 = No
Int
0
1
4
(0x04)
1
Dep/Etch Mode
0 = Deposit
1 = Etch
Int
0
1
5 (0x05)
1
Test
1 = Yes, 0 = No
Int
0
1