S12. . . . . Input status - returns 9 ASCII bytes in order byte 1 to 9, 1 byte per input.

Each byte has an ASCII value of 0 or 1, corresponding to the input’s
status. 0 = grounded (active), 1 = open or high (inactive).

S13. . . . . Raw frequency - Frequency of crystal being read. xxxxxxx.x Hz

[negative of last good frequency if failed]

S14. . . . . Xtal Fail - Returns ASCII 1 if currently failed crystal, 0 if not.

S15. . . . . Max Power - Returns ASCII 1 if currently outputting maximum power,

0 if not.

S16. . . . . Crystal switching - Returns ASCII 1 if currently crystal switching, 0 if not.

S17. . . . . End of process - Returns ASCII 1 if process has ended, 0 if not.

S18. . . . . STOP - Returns ASCII 1 if process is in STOP.

S19. . . . . DATALOG - Returns the datalog string, refer to

section 5.2.2.5

, S19

command

for details. Data is separated by spaces instead of CR/LF.

The last byte returned identifies the End on Time Power or Normal Completion
information as 1 or 0 respectively. Also, when using the S19 command the
Begin Process and End Process messages are not returned.

S20. . . . . Present Configuration Selections - returns 16 ASCII bytes with

a value of 0 or 1.

Data byte order (Non-checksum): 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 <ACK>

Table 5-10 S12 Command Input status

Input #

Byte #

Function

1

Start

2

Stop

3

End

4

Sample Initiate

5

Sample Inhibit

6

Crystal Fail Inhibit

7

Zero Thickness

8

Soak 2 Hold

9

Crucible Valid