3 - 19

PN

07

4-

44

6-

P1

J

XTC/3 Operating Manual

Table 3-2 Output states

Output

Active

Inactive

Source Shut 1

In any deposit state

When leaving control states

Source Shut 2

1

Sensor Shut 1

In shutter delay or any deposit
state.
In RateWatcher Sample

Closes when leaving control states.
In RateWatcher Hold

1

Sensor Shut 2

Note 1: In XTC/3S, Sensor Shut 1 and Sensor Shut 2 outputs are used to perform the crystal
switch function if a multi-position sensor has been selected.

Stop

Enter Stop

Exit Stop

End of Process

When entering Idle if this is last
film in the process

When Start or Reset is done

Thick Setpoint

In Deposit or Time Power state,
and thickness setpoint is reached

When Start or Reset is done
When Idle is entered

Xtal Fail

When active crystal fails

When active crystal becomes valid.
Digital Set Crystal Fail Inhibit

Alarms

On Rate Deviation error
On Shutter Delay error
When Max power reached
On Crucible Switch error

As each error is cleared and no other
error is set.
When Start or Reset is done
When Idle ramp is entered

Source in use

When Source 2 is active

When Source 1 is active

Final Thick

Final thickness is reached

When Start or Reset done

End of Film (M only)

Initiate Idle state

When Start or Reset done

In Layer (M only)

When layer is started

Enter Idle or Stop

Ion Assist Dep
(M only)

Enter any deposit state and Ion
Assist Deposit is enabled

Enter Idle ramp
Enter Stop

Xtal Switch 1
(M only)

See crystal switching

section 3.6.8

See crystal switching

section 3.6.8

Xtal Switch 2
(M only)

See crystal switching

section 3.6.8

See crystal switching

section 3.6.8

Advertising