Section 6.5.1.2 – INFICON XTC/3 Thin Film Deposition Controller Operating Manual User Manual

Page 188

Advertising
background image

6 - 26

PN

07

4-

44

6-

P1

J

XTC/3 Operating Manual

6.5.1.2 Feed-Through Or In-Vacuum Cable

Diagnostic Procedure

1

Refer to

Figure 6-10 on page 6-24

. Remove the In-Vacuum cable from the

Sensor Head at point B.

2

Connect the Crystal Sensor Emulator to the In-Vacuum cable.

If the XTAL Fail message disappears after approximately five seconds, the
Feed-Through and In-Vacuum Cable are working properly. Re-install the
In-Vacuum cable to the Sensor Head. Go to

section 6.5.1.3 on page 6-27

.

If the XTAL Fail message remains, continue at step 3.

3

Disconnect the In-Vacuum cable from the Feed-Through and the Emulator.
Disconnect the 6 in. BNC cable from the Feed-Through.

4

Using an Ohm Meter, verify electrical continuity from the BNC center pin on the
Feed-Through to the Microdot center pin on the Feed-Through. A typical value
would be less than 0.2 Ω.

5

Verify electrical isolation of the center pin on the Feed-Through from the
electrical ground (Feed-Through body). A typical value would be in excess of
10 MΩ.

If the Feed-Through is found to be defective, replace the Feed-Through, re-attach
the BNC and In-Vacuum cables, and repeat this procedure starting at Step 2,
otherwise continue at step 6.

6

Verify electrical continuity from center pin to center pin on the In-Vacuum cable.

7

Verify that the center pin of the In-Vacuum cable is electrically isolated from the
In-Vacuum cable shield.

If the In-Vacuum cable is found to be defective, replace the In-Vacuum cable.
Re-attach the BNC and In-Vacuum cables, and repeat this procedure starting at
Step 2, otherwise continue at step 8.

8

Connect the In-Vacuum Cable to the Feed-Through.

9

Verify electrical continuity from the center pin on the BNC connector of the
Feed-Through to the center pin on the un-terminated end of the In-Vacuum
cable.

10

Verify electrical isolation from the center pin to electrical ground (Feed-Through
body).

If the Feed-Through/In-Vacuum cable system is found to be defective, look for
defective electrical contacts at the Feed-Through to In-Vacuum cable connection.
Repair or replace the Feed-Through as necessary. Re-attach the BNC
and In-Vacuum cables and repeat this procedure starting at step 2.
Otherwise, continue at step 11.

Advertising