6.7 TEST Page Set Up

Figure 6-6 General test page

TEST . . . . . . . . . . . . . . . . . . . . . . . . On (1), Off (0)

Press the TOGL key to move through the choices. The default is Off. The
Cygnus 2 contains a software controlled Test Mode which simulates actual
operation. The purpose of the Test Mode is to verify basic operation and for
demonstrating typical operation to the technician. The word "TEST" will be
diaplayed when TEST is on.

The Rate displayed during Test Mode operation is determined as shown in

equation [1]

.

[1]

Relays, inputs, Source and recorder outputs operate normally during Test
Mode but Crystal fails are ignored unless Advanced Test is set to On.

TIME COMPRESSED . . . . . . . . . . . Yes (1), No (0)

When in TEST mode, TIME COMPRESSED allows a faster than real time
execution of the Process. This is useful when testing a long Process. A value
of Yes increases the speed of execution 10 to 1. The default value is No.

ADVANCED TEST . . . . . . . . . . . . . . On (1), Off (0)

If set to On, the Cygnus 2 will respond to crystal fails and switch to backup
crystal positions. The text "A TEST" will be displayed when the Advanced Test
is On. The default value is Off.

Displayed Rate

40

DENSITY (gm/cc)

----------------------------------------------- TOOLING (%)

100%

---------------------------------------

·

Å/sec

=