31 remote general action – INFICON Cygnus 2 Thin Film Deposition Controller User Manual

Page 190

Advertising
background image

9 - 42

IP

N 07

4-

54

5-

P1

E

Cygnus 2 Operating Manual

9.4.31 Remote General Action

Command = RG <Command ID> (<Action Value>)

Command ID = <Byte> See

Table 9-15

.

Action Value = <Byte>… <Byte> | <Integer>…<Integer>|<Float>…<Float>

See

Table 9-15

for commands that require an action value.

Response = None (Just header and trailer)

Table 9-15 Remote general

RG

Command

ID

Meaning

Action Value

Notes

1 (0x01)

Stop All

Duplicates the equivalent front panel action

2 (0x02)

Reset All

Duplicates the equivalent front panel action

3 (0x03)

Remote Lock On

No parameters can be updated from the front panel

"R Lock" displayed

4 (0x04)

Remote Lock Off

Cancel RG 3 command

5

(0x05)

Set Logic

Statement

<Statement

Number> to True

<Byte>

Range is 1-100.

Only does something if logic statement event "Computer

Control" set. Otherwise, ignored.

6

(0x06)

Clear Logic

Statement

<Statement

Number> to False.

<Byte>

Range is 1-100.

Only does something if logic statement event "Computer

Control" set. Otherwise, ignored.

7

(0x07)

Crystal Fail Inhibit

On

<Byte>

Range is 0-6 (0 = All)

"Crystal Fail" output always set to FALSE, even during

crystal fail

8

(0x08)

Crystal Fail Inhibit

Off

<Byte>

Range is 0-6 (0 = All)

"Crystal Fail" output has normal meaning

9

(0x09)

Set date (Month,

Day, Year) or (Day,

Month, Year)

Depending on

General parameter

17 (Date Format).

<Integer>
<Integer>
<Integer>

Range as follows:Month = 1 to 12

Day = 1 to 31

Year = 2000 to 2099

10

(0x0a)

Set time (Hour,

Minute, Second)

<Integer>
<Integer>
<Integer>

Range as follows Hour = 0 to 23

Minute = 0 to 59

Second = 0 to 59