9.4.31 Remote General Action
Command = RG <Command ID> (<Action Value>)
Command ID = <Byte> See
Action Value = <Byte>… <Byte> | <Integer>…<Integer>|<Float>…<Float>
See
for commands that require an action value.
Response = None (Just header and trailer)
Table 9-15 Remote general
RG
Command
ID
Meaning
Action Value
Notes
1 (0x01)
Stop All
Duplicates the equivalent front panel action
2 (0x02)
Reset All
Duplicates the equivalent front panel action
3 (0x03)
Remote Lock On
No parameters can be updated from the front panel
"R Lock" displayed
4 (0x04)
Remote Lock Off
Cancel RG 3 command
5
(0x05)
Set Logic
Statement
<Statement
Number> to True
<Byte>
Range is 1-100.
Only does something if logic statement event "Computer
Control" set. Otherwise, ignored.
6
(0x06)
Clear Logic
Statement
<Statement
Number> to False.
<Byte>
Range is 1-100.
Only does something if logic statement event "Computer
Control" set. Otherwise, ignored.
7
(0x07)
Crystal Fail Inhibit
On
<Byte>
Range is 0-6 (0 = All)
"Crystal Fail" output always set to FALSE, even during
crystal fail
8
(0x08)
Crystal Fail Inhibit
Off
<Byte>
Range is 0-6 (0 = All)
"Crystal Fail" output has normal meaning
9
(0x09)
Set date (Month,
Day, Year) or (Day,
Month, Year)
Depending on
General parameter
17 (Date Format).
<Integer>
<Integer>
<Integer>
Range as follows:Month = 1 to 12
Day = 1 to 31
Year = 2000 to 2099
10
(0x0a)
Set time (Hour,
Minute, Second)
<Integer>
<Integer>
<Integer>
Range as follows Hour = 0 to 23
Minute = 0 to 59
Second = 0 to 59