Cygnus 2 Operating Manual

9.4.30 Status Sensor

Command = SS <Command ID><Sensor>

Command ID = <Byte> See "Description" in

Table 9-14

.

Sensor = <Byte> Sensor number (0 - 6)

If the Sensor number is zero the response will give the value for all six Sensors.

Response = See Data Returned in

Table 9-14

.

Table 9-14 Status sensor

SS

Command

ID

Description

Number
of bytes

in the

response

Data Returned

Notes

0

(0x00)

Crystal Life

1 or 6

<Byte> or

<Byte><Byte><Byte>
<Byte><Byte><Byte>

One byte for each Sensor.

Range will be 0 - 100

1

(0x01)

Crystals

Remaining

1 or 6

<Byte> or

<Byte><Byte><Byte>
<Byte><Byte><Byte>

One byte for each Sensor. Range between 0

and 12. Number of good crystals on a

CrystalTwo, Crystal6 or Crystal12 Sensor.

Returns either a 0 (current crystal failed) or a

1 (current crystal good) for a Single sensor.

The command is not intended to be used for

Generic sensor - returns NO_DATA .

2

(0x02)

Crystal

Position

1 or 6

<Byte> or

<Byte><Byte><Byte>
<Byte><Byte><Byte>

One byte for each Sensor. Range between 1

and 12. The current crystal position on a

CrystalTwo, Crystal6 or Crystal12.

3

(0x03)

Sensor

Status

1 or 6

<Byte> or

<Byte><Byte><Byte>
<Byte><Byte><Byte>

One byte for each Sensor.

Bit1 Bit 0

0 0 Good Crystal
0 1 Failed Crystal
1 0 Invalid Measurement

on Crystal

1 1 Undefined
Bit 7 Bit 6
0 0 Auto-Z-Ratio
0 1 Sensor Z-Ratio
1 0 Material Z-Ratio
1 1 Undefined

4

(0x04)

Fundamental

Frequency

8 or 48

<Long> or

<Long><Long><Long>
<Long><Long><Long>

One long integer for each Sensor.

Unconverted. To convert to Hz, multiply by

0.000873114913702011