1 troubleshooting the cygnus 2 – INFICON Cygnus 2 Thin Film Deposition Controller User Manual

Page 228

Advertising
background image

13 - 12

IP

N 07

4-

54

5-

P1

E

Cygnus 2 Operating Manual

13.6.1 Troubleshooting the Cygnus 2

Table 13-6 Troubleshooting the Cygnus 2

SYMPTOM

CAUSE

REMEDY

1. Power on LED not
illuminated

a. Blown fuse/circuit breaker
tripped

a.Have qualified personnel
replace fuse/reset circuit
breaker

b. Electrical cord unplugged
from wall or back of Cygnus 2

b. Reconnect power cord

c. Incorrect line voltage

c. Have qualified personnel
verify line voltage, verify the
Cygnus 2 is configured for
the correct voltage

2. Cygnus 2 "locks" up

a. Cover or back panels not
attached to the Cygnus 2

a. Ensure all covers and
panels are in place and
securely fastened

b. High electrical noise
environment

b. Reroute cables to reduce
noise pickup (1 ft [30 cm]
away from high power
conducting lines makes a
sizable reduction in the
amount of noise entering the
Cygnus 2), keep all ground
wires short with large surface
area to minimize ground
impedance (solid copper
strap with 5 to 1 length to
width ratio is recommended)

c. Poor grounds or poor
grounding practice

c. Verify proper earth ground,
use appropriate ground
strap, (solid copper strap
with 5 to 1 length to width
ratio is recommended),
eliminate ground loops by
establishing the correct
system grounding, verify
proper Cygnus 2 grounding

3. Cygnus 2 does not retain
parameters on power down
(loss of parameters on power
up)

a. Lithium battery is dead

a. Lithium battery has a
normal life expectancy of
seven years, contact
INFICON service department

b. Power supply or Processor
board problem

b. Contact INFICON service
department

Advertising