5 update general parameter, Table 9-5 – INFICON Cygnus 2 Thin Film Deposition Controller User Manual

Page 160

Advertising
background image

9 - 12

IP

N 07

4-

54

5-

P1

E

Cygnus 2 Operating Manual

9.4.5 Update General Parameter

Command = UG <Command ID> <Parameter Value>

Command ID = <Byte> See

Table 9-5

Parameter Value = <Integer> | <Float> Type of value depends on command ID,
see "Data Type" column in

Table 9-5

.

Response = None (Just header and trailer)

General Parameters

Table 9-5 General parameters

QG

&

UG

Cmnd

ID

Name

(Units)

Range:

Low

Limit/

High

Limit/

Default

Data Type

Notes

Update Restrictions

and Display Notes

9

(0x09)

Thickness Eq 1

0/

654321/

0

Integer

Values are the Sources

whose thicknesses are

summed

No duplicate numbers

allowed

10

(0x0a)

Thickness Eq 2

11

(0x0b)

Thickness Eq 3

12

(0x0c)

Test

0/ 1/ 0

Encode

0 = Off
1 = On

13

(0x0d)

Time Compressed

(10x time

acceleration)

0/ 1/ 0

Encode

0 = No

1 = Yes

Gray font when Test = Off

14

(0x0e)

Advanced Test

(Crystal Fails

active)

0/ 1/ 0

Encode

0 = Off
1 = On

15

(0x0f)

Audio Feedback

0/ 1/ 0

Encode

0 = No

1 = Yes

16

(0x10)

LCD Dimmer Time

(Minutes)

0/ 99/ 0

Integer

0 = Always On

Non 0 = Minutes till off

17

(0x11)

Date Format

0/ 1/ 0

Encode

0 = mm/dd/yyyy
1 = dd/mm/yyyy

18

(0x12)

USB Datalog

Format

0/ 2/ 0

Encode

0 = Log Off 1 = Page

2 = Comma

Advertising