8.4 THEN Action Definitions

NOTE: The following are level outputs that once turned on will remain active until

turned off. This is true even if the Logic string is cleared.

START INHBT (ON/OFF)

XTL FL INHB (ON/OFF)

RWS INHIBIT (ON/OFF)

RWS INITIATE (ON/OFF)

SOAK 1 HOLD (ON/OFF) ##

SOAK 2 HOLD (ON/OFF) ##

EXTERN OUT (ON/OFF) ##

MESSAGE (ON/OFF) ##

CLOCK HOLD (ON/OFF)

PRE COND S/H (ON/OFF)

CANCEL TIMER ## . . . . . . . . . . . . . 1 to 20

Deactivates the designated timer ## and resets it to zero.

CLEAR COUNTER ## . . . . . . . . . . . 1 to 20

Clears the designated counter ##.

CLOCK HOLD (ON/OFF) ## . . . . . . 0 (any Material) to 6

This feature controls the state timer during any non-rate control state. These
states include the pre-deposit states—Ready, Crucible Switch, PreCond Rise,
PreCond Soak, Rise 1, Soak 1, Rise 2, Soak 2—and the post-deposit state Idle
Ramp.

For the feature to be activated the state’s timer must be non-zero. If the
Cygnus 2 is in the Ready state and a Start command is executed while CLOCK
HOLD is active, the Cygnus 2 will progress to the first pre-deposit state with a
non-zero state time. If the Cygnus 2 is in the Crucible Switch state, waiting for
Turret Feedback Input, and CLOCK HOLD is activated, when the turret
positioned input is activated, the Cygnus 2 will progress to the next pre-deposit
state having a non-zero state time. The Cygnus 2 is prevented from continuing
state processing until the CLOCK HOLD action is turned OFF.

CONTINUE DEP ## . . . . . . . . . . . . . 0 (any Material) to 6

Maintains the designated Material ## in the Deposit state but does not zero the
thickness. This action cannot be done from a Ready, Idle, Stop, Crystal
Switching, or Crucible Switch state. If the crystal is failed, this action will put the
associated Source into the Time Power state if the previous state was
Non-Deposit Hold, otherwise it is not allowed.

EXTERN OUT (ON/OFF) ## . . . . . . 1 to 38

Designates a hardware output ## and places it into its ON or OFF state.
Outputs can be given names up to 15 characters long. Outputs are named in
the I/O Map display.