14 query material name, 15 update material name, 16 query input name – INFICON Cygnus 2 Thin Film Deposition Controller User Manual
Page 173: 17 update input name, 18 query output name
9 - 25
IP
N 07
4-
54
5-
P1
E
Cygnus 2 Operating Manual
9.4.14 Query Material Name
Command = QN <Material Number>
Material Number = <Byte> 1-6
Response = < Material Name>
Material Name = <String> Up to 15 characters, null terminated. The string returned
will be a blank if the default Material # name has not been changed.
9.4.15 Update Material Name
Command = UN < Material Number> < Material Name>
Material Number = <Byte> 1 - 6
Material Name = <String> Up to 15 characters, null terminated
Response = None (Just header and trailer)
9.4.16 Query Input Name
Command = QI <Input Number>
Input Number = <Byte> 1 - 28
Response = <Input Name>
Input Name = <String> Up to 15 characters, null terminated
9.4.17 Update Input Name
Command = UI <Input Number> <Input Name>
Input Number = <Byte> 1 - 28
Input Name = <String> Up to 15 characters, null terminated
Response = None (Just header and trailer)
Input Names may only be updated if the Cygnus 2 is in Ready or Stop.
NOTE: UI changes are not allowed while a Material is running. The UI command
will not allow updating a name set by hardware designation, unless it is
identical to the defined hardware name.
9.4.18 Query Output Name
Command = QO <Output Number>
Output Number = <Byte> 1 - 38
Response = <Output Name>
Output Name = <String> Up to 15 characters, null terminated