6 lib a-hf, lib hf-sc, lib sc-z sub-displays, 7 general – INFICON Cygnus 2 Thin Film Deposition Controller User Manual

Page 67

Advertising
background image

3 - 17

IP

N 07

4-

54

5-

P1

E

Cygnus 2 Operating Manual

3.3.6.6 Lib A-Hf, Lib Hf-Sc, Lib Sc-Z Sub-Displays

These three libraries allow convenient selection of a Material by its chemical
formula along with the correct density and Z-Ratio values. Cursor to a Material
formula, then press F1 Define Material to select that Material. The Material/Source
display will be shown where the remaining parameters can be entered. See

section

5.1.2, Material Definition, on page 5-2

.

3.3.7 General

The General screen is organized into eight sub screens described below.

See

Chapter 6, General Parameters

for programming details and screen displays.

3.3.7.1 Process Page

The Process page shows global parameters and allows editing them.

Figure 3-12 Process page

Thickness Eq 1, 2 and 3. . . . . . . . . allows summing of thicknesses from up to 6

Sources for use in conjunction with a logic
statement Thick Sum event.

See

section 6.2, General Process Page, on page 6-1

3.3.7.2 DACs Page

Polarity and full scale voltage for all DAC outputs are configured here. 0 output
corresponds to 0 Volts. See

section 6.3, DACs Page Parameters, on page 6-2

.

Advertising