13.5 Input Error Messages

Keyboard Input Error Messages come up on the entry of invalid parameters. These
messages will appear on the transient error message line in the message box. The
font color of the parameter value being updated will change to a different color to
denote an input error has been made. They will clear when the clear key is pressed.
The original stored value will then be re-displayed. See

Table 13-5

.

Table 13-5 Input error messages

Input Error Message

Set

Cleared

Screen

Value Too High

The value entered is larger than the maximum
value allowed.

Each input error
message is
taken down by
pressing the
clear key.

Any Parameter
edit

Value Too Low

The value entered is lower than the minimum
value allowed.

Any Parameter
edit

Duplicate Source

Can use each Source number only once for
thickness sum.

General

Invalid Mask

An attempt was made to enter an incorrect
Ethernet net mask.

General

Incorrect Time

An attempt was made to enter more then 59
seconds or 99 minutes into a time parameter.

Material

Ramp 1 > Ramp 2

Invalid Material parameter ‘Ramp’, values were
entered. Start Ramp 1 thickness value must be
less then Start Ramp 2 value when the Start
Ramp 2 value is non-zero.

Material

Ramp 2 < Ramp 1

Invalid Material parameter ‘Ramp’, values were
entered. Start Ramp 2 thickness value must be
greater then Start Ramp 1 value.

Material

DAC Used Material ##
Or
DAC Used Sensor ##

The DAC number has been previously defined
as a recorder output in the Material or the Sensor
set up.

Source
parameter

DAC Used Source XX

The DAC number has been previously defined
as a Source DAC output in one of the Source set
ups

Material, Sensor
and Source
parameter

Input in use

An attempt was made to enter an input that is
already reserved for use in Source setup or one
of the logic statements.

Source
parameter

Output in use

An attempt was made to enter an output that is
already reserved for use. The output is already
reserved for use in Sensor or Source set up or in
one of the logic statements.

Sensor, Source
parameter and
Logic Edit

Invalid Lock Code

An incorrect lock code was used to attempt to
unlock the Cygnus 2.

Menu