3 shuttered and dual sensors, 4 bakeable sensor – INFICON Cygnus 2 Thin Film Deposition Controller User Manual

Page 239

Advertising
background image

13 - 23

IP

N 07

4-

54

5-

P1

E

Cygnus 2 Operating Manual

13.7.3 Shuttered and Dual Sensors

There is no difference in the crystal replacement procedure between shuttered and
non-shuttered Front Load sensors, since the shutter pivots away from the crystal
opening when the shutter is in the relaxed state. This is also the case for Cool
Drawer Sensors as the shutter assembly does not obstruct the drawer from being
inserted into the sensor body.

13.7.4 Bakeable Sensor

For the Bakeable sensor, the procedure is the same as the Front Load sensor
except that you must first unlock the cam assembly by flipping it up. Once the
crystal has been replaced, place a flat edge of the holder flush with the cam
mechanism and lock it in place with the cam. See

Figure 13-4

.

Figure 13-4 Bakeable crystal sensor

Crystal Holder and

Retainer Spring

IPN 750-218-G1

Crystal IPN 008-010-G10

Fully Coated Face (Gold)

Crystal Retainer IPN 007-064

Clamping Spring IPN 007-094

Contact IPN 007-099

Insulator IPN 007-103

Spreader Bar
IPN 007-267-P2

Shoulder Washer

IPN 007-268-P1
Two Required

Shoulder Washer

IPN 007-269-P1

Cam Mechanism

IPN 007-095

Cover IPN 007-101

Advertising