Chapter 2

Installation and Interfaces

2.1 Location Guidelines

Before permanently installing the Cygnus 2, read this entire chapter on installation
and interfaces and follow its recommendations as closely as possible. INFICON
has taken numerous steps to ensure its equipment will operate in a variety of harsh
situations. Failure to adhere to these simple practices may adversely affect the
performance and longevity of the Cygnus 2.

2.1.1 Sensor Types

The choice of sensor type must be dictated by the process, the deposition material
and the physical characteristics of the process chamber. General guidelines for
each sensor type produced by INFICON are outlined in the Sensor Data Sheets on
the

www.inficon.com

website. For specific recommendations, consult your

INFICON representative.

CAUTION

The performance of this Cygnus 2 depends on the careful
installation of the chosen transducer. Improper
installation will cause problems with deposition
repeatability, crystal life and rate stability.

2.1.2 Sensor Installation

Figure 2-1

shows a typical installation of an INFICON water cooled crystal sensor

in the vacuum process chamber. Use the illustration and the following guidelines to
install your sensors for optimum performance and convenience.