Metrohm tiamo 1.1 Manual User Manual

Page 147

Advertising
background image

Subwindow Run

119

not yet been assigned in a command then the window

Device assign-

ment

appears for assigning the device.

• Check solutions

Checks whether the solutions demanded in the device-specific com-
mands are present and available.

• Check sensors

Checks whether the sensors demanded in the device-specific com-

mands are present and available.

Note
The Start test can also be triggered manually in the

READY

status with the menu

item

Tools, Run test

.

4 - Preconditioning (only for methods with conditioning)

In methods that contain commands with switched-on conditioning and in which
the option

Automatic conditioning

is switched on in the

START

command then pre-

conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (

Sample number

,

Statis-

tics

).

• Conditioning requirement not fulfilled

After the start of preconditioning the status changes to

COND BUSY

. In

this status conditioning can be started with

[Start]

even though the

conditioning requirement has not been fulfilled. Conditioning can also
be canceled with

[Stop]

or interrupted with

[Hold]

and then continued

with

[Cont]

. If an error occurs during conditioning then neither the Error

track nor the Exit track will be carried out.

• Conditioning requirement fulfilled

If the conditioning requirement has been fulfilled then the status
changes to

COND READY

. In this status the determination can be

started with

[Start]

. Conditioning can also be canceled with

[Stop]

or

interrupted with

[Hold]

and then continued with

[Cont]

. The condition-

ing requirement will continue to be monitored continuously until the
determination has been started.

5 - Start method

At the start of the method the following checks and actions are triggered:

• Check sample data

Checks that the entered Sample data corresponds to that defined in

the Method variables and is valid.

• Assign determination ID

An unambiguous determination ID is assigned for the determination.

• Reserve device (module)

The devices (or their functional units) used by the device-dependent

commands contained in the method are reserved for the duration of
the determination, i.e. they cannot be used in other workplaces nor
can they be operated manually.

• Increment sample number

The

Sample number

is increased by

+1

.

• Increment statistics counter

If statistics has been switched on in both the

START

command and on

the tab

Single determinations

then the statistics counter is increased by

+1

. If a new method is loaded or if the statistics counter agrees with