Method

Evaluation with measured value window (pH)

on, [ off ]

With this option set up to nine regions (windows) can be defined on the
measured value axis. Only those equivalence points which lie within these

windows and additionally meet the parameters defined for each window will
be recognized. Per window only one EP will be recognized. The defined
windows with their parameters are shown in the table and can be edited
with the following buttons:

Opens the dialog window

Measured value window #

(see below) in which the

parameters for a new window can be entered.

Opens the dialog window

Measured value window #

(see below) in which the

parameters for a new window can be edited.

Deletes the selected window in the table.

EP criterion

0 ... [ 5 ] ... 200

(only Titrino)

Criterion for the recognition of equivalence points which is valid for all win-

dows. Equivalence points whose ERC is smaller than the set EP criterion
will not be recognized.

Measured value window (pH)

Measured value windows are regions on the measured value axis for which differ-
ent parameters for the potentiometric evaluation can be defined. Only those
equivalence points which lie within these windows and additionally meet the pa-
rameters defined for each window will be recognized. Per window only one EP will

be recognized.