Additional evaluations – Metrohm tiamo 1.1 Manual User Manual

Page 365

Advertising
background image

Commands

337

Additional evaluations

The following additional methods for evaluation of titration curves can be activated

and defined here:

Command name

25 characters, [ DET pH # ]

Name of the command.

Fix end point evaluation

on, [ off ]

For a fixed quantity (measured value, volume or time) for the fix end point

the associated values are interpolated from the list of measuring points.

Fixed quantity

[ Measured value ], Time, Volume

Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measured points.

Fix EP# at pH

-20.000 ... 20.000, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Measured value

.

Fix EP# at

0.0 ... 999999.9 s, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Time

.

Fix EP# at

0.00000 ... 9999.99 mL, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Volume

.

With this button additionally to the always shown fix EP1...3 the fix EP4...9
are displayed.

With this button instead of fix EP1...9 only the fix EP1...3 are shown.