Metrohm tiamo 1.1 Manual User Manual

Page 504

Advertising
background image

476

Method

Min. waiting time

[ 0 ] ... 999'999 s

(Titrando only)

The measured value is only accepted if the minimum waiting time has
elapsed even if the signal drift has already been reached. The drift contin-
ues to be checked as the minimum waiting time is passing.

Max. waiting time

0 ... [ 52 ] ... 999'999 s

(Titrando)

0 ... [ 52 ] ... 9'999 s

(Titrino)

If the signal drift has not yet been reached then the measured value will be
accepted when the maximum waiting time has elapsed. If no new entry for
the waiting time has been made then a suitable value for the drift will be
calculated automatically according to the following equation:

Measuring interval

0.1 ... [ 2.0 ] ... 999'999.0 s

(Titrando, increment = 0.1 s)

0.08 ... [ 2.0 ] ... 16'200 s

(Titrino, increment = 0.08 s)

Time interval for entering a measuring point in the list of measuring points.

Stop measured value

-20.0 .. 150.0 °C, [ off ]

(Titrando only)

Stops when the preset measured value has been reached since the start of
the measurement.

off

means no stop.

Measurement without drift control

[ on ], off

The measurement is carried out without drift control if this option is acti-
vated. The measurement continues until one of the stop criterions

Measur-

ing time

or

Stop measured value

is met. The corresponding parameters are

not visible if this option is deactivated.

Measuring time

0 ... [ 120 ] ... 999'999 s

(Titrando)

0 ... [ 120 ] ... 9'999 s

(Titrino)

Maximum period of time for measurement.

Measuring interval

0.1 ... [ 2.0 ] ... 999'999.0 s

(Titrando, increment = 0.1 s)

0.08 ... [ 2.0 ] ... 16'200 s

(Titrino, increment = 0.08 s)

Time interval for entering a measuring point in the list of measuring points.

Stop measured value

-20.0 ... 150.0 °C, [ off ]

(Titrando only)

Stop when the preset measured value has been reached since the start of
the measurement.

off

means no stop.