Metrohm tiamo 1.1 Manual User Manual

Page 427

Advertising
background image

Commands

399

displayed. In order to edit these parameters the value

user

must be se-

lected.

slow

For titrations in which the finest details are also to be visible. However, this
can also increase noise and lead to the production of unwanted EPs.

[ optimal ]

Parameter set for all standard titrations; optimized for the most frequent
applications.

fast

For less critical rapid titrations.

user

Editing the individual titration parameters which affect the titration rate.

Control

(only visible for

Titration rate

=

user

)

Dynamics

0.01 ... [ 40.00 (slow) ] ... [ 10.00 (optimal) ] ... [ 5.00 (fast) ] ... 200.00 µA, off

(Titrando)

0.1 ... [ 40.0 (slow) ] ... [ 10.0 (optimal) ] ... [ 5.0 (fast) ] ... 200.0 µA, off

(Titrino)

Dynamics defines the measured value range before the given endpoint.
This parameter has an important influence on the titration rate and with it on
the accuracy. Within the control range the steps which are added are con-

trolled by the

Min. rate

. The nearer the endpoint the slower the dosing takes

place until the

Min. rate

has been reached. Outside the range the dosing

takes place with

Max. rate

(see Control range).

Max. rate

0.01 ... [ 1.00 (slow) ] ... [ 10.00 (optimal) ] ... 166.00 mL/min, [ maximal (fast) ]

(Titrando)

0.01 ... [ 1.00 (slow) ] ... [ 10.00 (optimal) ] ... 150.00 mL/min, [ maximal (fast) ]

(Titrino)
Speed of addition outside the control range. The maximum dosing speed
depends on the cylinder volume of the exchange unit or dosing unit used.

Min. rate

0.01 ... [ 5.00 (slow)] ... [ 25.00 (optimal) ] ... [ 50.00 (fast) ] ... 9999.00 µL/min

Speed of addition at the start and finish of the titration. The slower the

minimum rate that is selected, the slower the titration and the higher the ac-
curacy.

Stop criterion

Stop criterion

[ drift ], time, off

The titration stops when the endpoint has been reached and the stop crite-
rion has been fulfilled. The titration can be stopped when a particular

drift

has been achieved or after a preset

time

. If no stop criterion has been se-

lected the titration will not be stopped. In older instructions the

Delay time

was usually defined as the

Stop criterion

. The stop criterion

time

means that

the endpoint must be exceeded for a certain time: the

Delay time

. The same

Delay time

with different very small volume increments (depending on the

volume of the exchange or dosing unit) means different switch-off points. In
contrast, if

drift

is used as the stop criterion then the switch-off point will al-

ways have the same curve slope dV/dt.

Note

The Stop conditions will always stop the titration even if the stop criterion has not
been fulfilled or has been turned off.

Advertising