Metrohm tiamo 1.1 Manual User Manual

Page 754

Advertising
background image

726

Configuration

Parameters for preparation

Configures the parameters to be used in the commands

PREP

and

EMPTY

.

Dosing port Prep/Empty

[ Dosing port 1 ], Dosing port 2, Fill port, Special port

Dosing port via which the cylinder contents are to be ejected during prepa-
ration and emptying.

Dosing rate Dosing port 1

0.01...166.00 mL/min, [ maximum ]

Speed at which dosing is to be carried out at Dosing port 1. The maximum
dosing rate depends on the cylinder volume of the dosing unit used (see

below). When the function is carried out the speed will automatically be re-
duced to the largest possible value.

Dosing rate Dosing port 2

0.01...166.00 mL/min, [ maximum ]

Speed at which dosing is to be carried out at Dosing port 2. The maximum
dosing rate depends on the cylinder volume of the dosing unit used (see

below). When the function is carried out the speed will automatically be re-
duced to the largest possible value.

Dosing rate Fill port

0.01...166.00 mL/min, [ maximum ]

Speed at which dosing or filling is to be carried out at the Fill port. The
maximum dosing rate depends on the cylinder volume of the dosing unit

used (see below). When the function is carried out the speed will automati-
cally be reduced to the largest possible value.

Dosing rate Special port

0.01...166.00 mL/min, [ maximum ]

Speed at which dosing is to be carried out at the Special port. The maxi-
mum dosing rate depends on the cylinder volume of the dosing unit used

(see below). When the function is carried out the speed will automatically
be reduced to the largest possible value.

Maximum Dosing rate for dosing unit as a function of the cylinder volume:

Cylinder volume

Maximum rate

2 mL

6.6 mL/min

5 mL

16.6 mL/min

10 mL

33.3 mL/min

20 mL

66.6 mL/min

50 mL

166.0 mL/min

Note
Enter lower rates for high-viscosity liquids.

Advertising