536

Method

Target

Tower

[ 1 ], 2

Selecting the tower of the Sample Processor for moving to the required po-

sition. Both Tower 1 and Tower 2 are always selectable even if the Sample
Processor has got only one tower.

Move

Selecting the target position on the rack:

[ Sample position ]

Position defined in the Run window as parameter

Sample position

.

Rack position

Position on the rack.

Special beaker

Special position on the rack defined in the rack properties (see Rack prop-
erties - Special beaker).

Relative angle

(only for 778, 789, 814, 815)

The sample rack can be rotated by a certain amount relatively to the current
position independently from rack positions. This can be used for example
to remove vial caps automatically.

Number

[ 1 ] ... 999

(for

Move

=

Rack position

)

[ 1 ] ... 16

(for

Move

=

Special beaker

, only for 778, 789, 814, 815, 855)

[ 1 ] ... 8

(for

Move

=

Special beaker

, only for 730, 774)Specification of the

number of the rack position. This parameter is only visible for

Move

=

Rack

position

or

Special beaker

. The rack position can be entered as a formula as

well. This is necessary for example for the automatic calibration of elec-

trodes. The Formula editor (see Formula editor) is opened by right clicking
into the input field.

Move angle

-259.90 ... [ 5.00 ] ... 259.90 °

This parameter specifies the angle about which the sample rack should be
rotated relatively to the current position. It is only visible for

Move

=

Relative

angle

.

Beaker test

Selection of the action to be carried out if the

Beaker sensor

defined for the

rack doesn't detect a sample at the target position. If the

Robotic arm

is de-

fined as beaker sensor, a Swing Head must be installed and a suitable

Work position

with beaker contact must be defined for the lift. This position

will be moved to automatically for beaker recognition after the command

MOVE

. This section is not visible for

Move

=

Relative angle

.

Display message

[ on ], off

A message is displayed and the determination is stopped.

Stop determination

on, [ off ]

The determination is stopped and the next determination of the series is
started.

Stop determination and series

on, [ off ]

The determination as well as the series are stopped. An Error track will be
started if there is one defined.

Parameters

Shift rate

3 ... [ 20 ] °/s

Entry of the speed at which the sample rack rotates.

Advertising