Commands

329

Start volume

Start volume

[ 0.00000 ] ... 9999.99 mL

(Titrando)

[ 0.000 ] ... 999.99 mL

(Titrino)

Volume to be added before the start of the titration with the below entered
speed.

Dosing rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando)

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the used exchange unit or dosing unit.

Start measured value

Start measured value pH

-20.000 ... 20.000, [ off ]

(only Titrando)

When the start measured value is reached the dosing of the start volume is

stopped and the next start condition is processed or the titration is started.
If the start measured value is achieved by the addition of a start volume
then the titration starts directly.

Dosing rate

0.01 ... [ 5.00 ] ... 166.00 mL/min, maximum

(only Titrando)

Speed at which the start volume is added until the start measured value is
reached. The maximum dosing rate depends on the cylinder volume of the
used exchange unit or dosing unit.

Start slope

Start slope

0.000 ... 9.999 pH/mL, [ off ]

(only Titrando)

When the start slope is reached the dosing of the start volume is stopped
and the titration is started. If the start slope is achieved by the addition of a
start volume then the titration starts directly.

Dosing rate

0.01 ... [ 5.00 ] ... 166.00 mL/min, maximum

(only Titrando)

Speed at which the start volume is added until the start measured value is

reached. The maximum dosing rate depends on the cylinder volume of the
used exchange unit or dosing unit.

Pause

Pause

[ 0 ] ... 999999 s

Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the addition of the start volume. The pause follows at the end

of all the start conditions.