Commands

445

Upper hysteresis

0.00 ... [ 0.20 ] ... 166.00 mL/min

(only Titrando)

The upper hysteresis represents a tolerance range for the upper limit of the
dosing rate. When the dosing rate falls below the upper limit plus this hys-
teresis value, the action

Dosing rate upper limit ok

is triggered.

Action

Definition of the action to be taken when the limits of the measured value
are infringed:

Stop determination

(only Titrando)

The running

STAT

command is quit, then the exit track (if present) is started

and the determination is stopped.

Stop command

The running

STAT

command is quit, then the next command is started.

Wait for [Continue]

Reagent dosing in the current

STAT

command is interrupted and a message

appears. As soon as the monitored dosing rate is again within the limits (in-
cluding hysteresis) reagent dosing can be resumed by pressing the

[Con-

tinue]

button in this message window.

Wait for limit ok

Reagent dosing in the current

STAT

command is interrupted. As soon as the

monitored dosing rate is again within the limits (including hysteresis) re-
agent dosing will be resumed automatically.

[ none ]

No action will be taken.

Monitoring Temperature

on, [ off ]

If this option is switched on then the temperature will be monitored and any
limit infringements will be entered in the list of measuring points.

Lower limit

[ -20.0 ] ... 150.0 °C

(Titrando)

[ -170.0 ] ... 500.0 °C

(Titrino)

Lower limit of the temperature. When the temperature falls below this value,

the action

Temperature lower limit exceeded

is triggered.

Lower hysteresis

0.0 ... [ 0.2 ] ... 150.0 °C

(only Titrando)

The lower hysteresis represents a tolerance range for the lower limit of the
temperature. When the temperature exceeds the lower limit plus this hys-

teresis value, the action

Temperature lower limit ok

is triggered

Upper limit

-20.0 ... [ 150.0 ] °C

(Titrando)

-170.0 ... [ 500.0 ] °C

(Titrino)

Upper limit of the temperature. When the temperature exceeds this value,
the action

Temperature upper limit exceeded

is triggered.

Upper hysteresis

0.0 ... [ 0.2 ] ... 150.0 °C

(only Titrando)

The upper hysteresis represents a tolerance range for the upper limit of the
temperature. When the temperature falls below the upper limit plus this hys-
teresis value, the action

Temperature upper limit ok

is triggered

Action

Definition of the action to be taken when the limits of the temperature are in-
fringed:

Stop determination

(only Titrando)

The running

STAT

command is quit, then the exit track (if present) is started

and the determination is stopped.

Stop command

The running

STAT

command is quit, then the next command is started.