Metrohm tiamo 1.1 Manual User Manual

Page 751

Advertising
background image

Subwindow Titrants/Solutions

723

Parameters for preparation

Configures the parameters to be used in the command

PREP

.

Volume

0.00000 ... 99999.99999 mL, [ cylinder volume ]

Volume to be dosed in during preparation. With

Cylinder volume

the whole

contents of the cylinder will be dosed in.

Cycles

1 ... [ 2 ] ... 9

Number of rinsing cycles during preparation.

Dosing rate

0.01 ... 150.00 mL/min, [ maximum ]

Speed at which the solution is to be dosed. The maximum dosing rate de-
pends on the cylinder volume of the exchange unit used (see below). When
the function is carried out the speed will automatically be reduced to the
largest possible value.

Filling rate

0.01 ... 150.00 mL/min, [ maximum ]

Speed at which the cylinder is to be filled. The maximum filling rate de-

pends on the cylinder volume of the exchange unit used (see below). When
the function is carried out the speed will automatically be reduced to the
largest possible value.

Maximum Dosing/Filling rate for exchange unit as a function of the cylinder volume:

Cylinder volume

Maximum rate

1 mL

3.0 mL/min

5 mL

15.0 mL/min

10 mL

30.0 mL/min

20 mL

60.0 mL/min

50 mL

150.0 mL/min

Note
Enter lower rates for high-viscosity liquids.

Tubing parameters

Defines the length and diameter of the tubing connected to the exchange unit.

Note
Default values have already been entered for the tubing parameters; these cor-
respond to the dimensions of the standard tubing supplied. If you do not make

any alterations to the tubing connections then you do not need to alter the tubing
parameters.

Dosing tip

Length

0.0 ... [ 40.0 ] ... 999.9 cm

Length of the tubing connecting the cock to the dosing tip.

Diameter

0.0 ... [ 2.0 ] ... 9.9 mm

Diameter of the tubing connecting the cock to the dosing tip.

Cylinder

Length

0.0 ... [ 13.0 ] ... 999.9 cm

Length of the tubing connecting the cock to the cylinder.

Advertising