494

Method

If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-

vices in the device table.

Dosing device

Dosing device

[ 1 ] ... 4

(808, 809, 835, 836, 842)

[ 1 ] ... 3

(855)

Selection of the dosing device (exchange or dosing unit) to be used for

dosing the addition solution.

Solution

24 characters, solution name, [ not defined ]

Input or selection of the solution defined in the solution table. Additionally,

not defined

can be selected. If intelligent exchange or dosing units are used

then a check will be made in the method sequence that the correct solution

is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder
volume is checked. At the start of the command the validity of the titer and
the expiry date of the selected solution will be checked as well as the GLP
test interval for the buret. If

not defined

is selected no tests will be carried

out.

Note
If an existing solution is selected as the solution to be added then the

Concentra-

tion

and

Unit

of this solution will be adopted for calculating the standard addi-

tion. If

not defined

is selected then

Concentration

and

Unit

must be entered on

the

Standard addition

tab.

Sensor

Measuring input

[ 1 ] ... 2

(Titrando)

[ 1 ]

(855)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensorname, [ ISE electrode (Fluoride) ]

Selection of the ISE sensor from the list of sensors defined in the sensor ta-

ble.

Temperature measurement

Type of temperature measuring:

continuous

A temperature sensor must be connected. The temperature is measured
continuously.

[ automatic ]

If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered

Temperature

(see

Meas-

uring parameters

) will be used.

off

The temperature will not be measured. The

Temperature

(see

Measuring pa-

rameters

) which has been entered manually will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

Setting the stirrer speed. The direction in which stirring takes place changes

as the sign in front of the stirrer speed changes.