Subwindow Sensors

735

Device

Select device, [ 'empty' ]

It is possible to select the device to which the sensor is connected. All the

devices in the device table which have a measuring input are shown in the
selection list. The connection of a sensor to a particular device is not
checked during the run.

Measuring input

[ 1 ], 2

It is possible to select the measuring input to which the sensor is con-

nected. The connection of a sensor to a particular measuring input is not
checked during the run. This parameter only appears if a

Device

has been

chosen.

Comment

24 characters

It is possible to enter remarks about the sensor.

Set to work

Select date

The current date is automatically entered here for a newly added sensor.
When editing an existing sensor the date can be selected by pressing

in the dialog window

Select date

(see General program functions - Edit - Se-

lect date).

Sensor monitoring

on, [ off ]

If this option is switched on then the working life of the sensor will be moni-
tored.

Working life

0 ... [ 999 ] days

Working life of the sensor in days. If a value is entered here then the

Expiry

date

will be adapted automatically.

Expiry date

Select date, [ Set to work + 999 days ]

Expiry date of the sensor. This date can be selected by pressing

in the

dialog window

Select date

(see General program functions - Edit - Select

date). If a date is entered here then the value for the

Working life

will be

adapted automatically.

Action

If during sensor monitoring it is found that the working life has expired then one of
the following actions will be triggered automatically at Start test:

Record message

The message that the working life of the sensor has expired will be auto-
matically saved with the determination.

Display message

A message will appear and you can choose whether to continue with the
run or cancel it. If the run is continued then the message that the working
life of the sensor has expired will be saved automatically with the determi-
nation.

Stop determination

The running determination will be terminated automatically. The following

message must be confirmed with

[OK]

.