Subwindow Sensors

737

concentration calibration is carried out with less than 3 standards then

c

(blank)

will be set to zero. The parameter

c (blank)

of an ISE electrode that

is used by the commands

MEAS Conc

and

CAL MEAS Conc

is available as a

variable

'Command name.BLV'

for calculations.

Calibration temperature

-20.0 ... [ 25.0 ] ... 150.0 °C

Temperature during the calibration. Whether the temperature was meas-
ured with a Pt1000 or an NTC temperature sensor, or entered manually, is

also shown.

Calibration date

Date and time of last calibration that is entered automatically after each
automatic calibration or manual entry.

Calibration method

Name of the method with which the last calibration was carried out. If the

calibration data was entered manually then

manual

will appear here.

User

Short name of the user logged in during the calibration or when the calibra-
tion data was entered manually. If work is not carried out using log in then
the user logged in under Windows will be entered automatically.

Calibration data monitoring

on, [ off ]

If this option is switched on then the validity of the calibration will be moni-
tored.

Calibration interval

0 ... [ 999 ] days

Validity of the calibration in days. If a value is entered here then the date for
the

Next calibration

will be adapted automatically.

Next calibration

Select date, [ Calibration date + 999 days ]

Date on which the next calibration is to be carried out. This date can be se-
lected by pressing

in the dialog window

Select date

(see General pro-

gram functions - Edit - Select date). When a date has been entered the
value for the

Calibration interval

will be adapted automatically.

Action

If during calibration monitoring it is found that its validity has expired then one of
the following actions will be triggered automatically at Start test:

Record message

The message that the validity of the calibration has expired will be auto-
matically saved with the determination.

Display message

A message will appear and you can choose whether to continue with the
run or cancel it. If the run is continued then the message that the validity of
the calibration has expired will be saved automatically with the determina-
tion.

Stop determination

The running determination will be terminated automatically. The following

message must be confirmed with

[OK]

.