Stop conditions, Conditioning – Metrohm tiamo 1.1 Manual User Manual
Page 445

Commands
417
Temperature
-170.0 ... [ 25.0 ] ... 500.0 °C
Manually entered titration temperature.
Time interval measuring point
[ 1 ] ... 999999 s
Time interval for entering a measuring point in the list of measuring points.
Stop conditions
Condition for stopping the titration.
Command name
25 characters, [ KFC # ]
Name of the command.
Stop time
0 ... 999999 s, [ off ]
Stops when the preset time has elapsed since the start of the titration.
off
means no stop.
Conditioning
Command name
25 characters, [ KFC # ]
Name of the command.
Conditioning
[ on ], off
Switching conditioning on/off. If conditioning is deactivated all following pa-
rameters are not visible.
Start drift
1 ... [ 20 ] ... 999 µg/min
The determination can only be started when the measured drift is below the
entered value. The Status in the
Run
window (Single determination or De-
termination series) then switches to
COND READY
.