131

determination will be terminated, if an error occurs then the Error track
starts.

7 - Post-conditioning (only for methods with conditioning)

In methods that contain commands with switched-on conditioning and in which
the option

Automatic conditioning

is switched on in the

START

command then for all

these commands post-conditioning will be started automatically with the loaded
method immediately after the end of the track that contains the conditioning

command. After the start of post-conditioning the status changes to

COND BUSY

.

In this status the conditioning can be stopped with

[Stop]

or interrupted with

[Hold]

and continued with

[Cont]

. If an error occurs during conditioning then neither the

Error track nor the Exit track will be carried out.

Conditioning between two determinations

Conditioning will continue to run until the new method is loaded for the
next determination.

Conditioning at the end of the last determination

After the Series end track has been carried out conditioning will con-
tinue to take place until it is terminated with

[Stop]

.

Device assignment

If at the Start test it is discovered that no device has been assigned for a device-
specific command then the window

Device assignment

appears for assigning the

device:

Command type

Shows the command type.

Command name

Shows the command name.

Device type

Shows the device type selected in the command.

Device name

Device name

Selection of a device from the device table. Only those devices being able
to execute the command are selectable.